Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Sub HP 545xx (1 Yr) Oscillator Output DATE: 08-Sep-98 AUTHOR: Fluke REVISION: $Revision: 1.1 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 1 NUMBER OF LINES: 56 ============================================================================= # # Source: # Hewlett Packard publication 54542-97045 # # Compatibility: # MET/CAL 5.1 or later # # Subprocedures: # Sub Wait for Operation Complete (OPC) IEEE # # Required Files: # HP545XX8.BMP # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 RSLT = 1.002 HEAD {OUTPUT SIGNAL CHECK - (Functional Check Only)} 1.003 TARGET 1.004 DISP This test checks the output of the AC Cal Output. 1.004 DISP THIS OUTPUT IS NOT SPEC'D and is entered as a 1.004 DISP functional check only. 1.004 DISP 1.005 PIC HP545XX8 1.006 IEEE :AUTOSCALE 1.007 CALL Sub Wait for Operation Complete (OPC) IEEE 1.008 JMPL END M[89] <= 0 # Set up measurement mode 1.009 IEEE :MEASURE:MODE STANDARD 1.010 IEEE :MEASURE:VAMPLITUDE 1.011 CALL Sub Wait for Operation Complete (OPC) IEEE 1.012 JMPL END M[89] <= 0 1.013 IEEE :MEASURE:VAMPLITUDE?[I] 1.014 MATH MEM2 = FMT(MEM, "%5.3f") 1.015 RSLT =Amplitude should be approximately 0.800V 1.016 RSLT =Actual value is [MEM2] 1.017 RSLT = 1.018 IEEE :MEASURE:SCRATCH 1.019 IEEE :MEASURE:FREQUENCY 1.020 CALL Sub Wait for Operation Complete (OPC) IEEE 1.021 JMPL END M[89] <= 0 1.022 IEEE :MEASURE:FREQUENCY?[I] 1.023 MATH MEM2 = FMT(MEM, "%5.3f") 1.024 RSLT =Frequency should be approximately 500Hz 1.025 RSLT =Actual value is [MEM2] 1.026 LABEL END 1.027 END